| Model Number |
03805002 |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Codes |
Injury (2348); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/23/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient was initially implanted on (b)(6) 2016 for l3-l4.On (b)(6) 2016, extension was performed for th7.Upon follow up, on (b)(6) 2018, it was confirmed a rod dislodged from the rod to rod clamp connector.There was no plan for revision surgery.There are no reports of adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received of patient experiencing pain and subsequently undergoing revision surgery of (b)(6) 2018.Section were updated accordingly.Method: x-rays, device history review, complaint history review, risk assessment; result: the customer reported event was confirmed via communication with the sales rep and x- ray images provided.The device was discarded at the hospital and is unavailable for evaluation.Additionally, no lot number was provided, so a device history review could not be performed.Conclusion: the x-ray shows that both rods slipped out of the rod to rod connectors.The rod length does not seem adequate in the x ray provided as there is not enough overhang to properly seat the rod into the connector and tighten down with the blockers.Additional factors that could have contributed to the event include patient post op activity, inadequate tightening of the connector blockers, excessive force applied, and patient factors.As the device was not returned for evaluation, a definite root cause cannot be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient was initially implanted on (b)(6) 2016 for l3-l4.On (b)(6) 2016, extension was performed for th7.Upon follow up, on (b)(6) 2018, it was confirmed a rod dislodged from the rod to rod clamp connector.Revision was performed (b)(6) 2018 due to patient pain.
|
| |
|
Search Alerts/Recalls
|