BARD ACCESS SYSTEMS GROSHONG 3F SINGLE-LUMEN PICC BASIC TRAY WITH MICROINTRODUCER; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
| Model Number |
N/A |
| Health Effect - Clinical Codes |
Edema (1820); Extravasation (1842); Pain (1994); Thrombosis (2100); Fibrosis (3167)
|
| Date of Event |
02/10/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned, at this time, to the manufacturer for evaluation.A lot history review (lhr) of reau1602 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (reau1602) have been reported from the same brazil facility.Device has not yet been returned for evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that patient using picc in the upper left limb showed pain and an increase of the brachial perimeter, edema in the upper left limb.After that the use of the picc was suspended, the case was assessed and measured and reassessed the next day it shows fibrous cord.It was also found that picc has a hole and a solution extravasation.On (b)(6) 3018 - it was reported that the "fibrous cord" is a consequence of a superficial thrombosis.It was stated that "the thrombotic process that progressed to fibrosis, which determines the permanence of the hardened venous cord.¿.
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a leak is unconfirmed, problem could not be reproduced.One 3 fr groshong catheter was returned for investigation.The catheter extended past the 32 cm depth marker.The sample was flushed with water using a 12 ml syringe and was found to be patent to infusion.The sample was pressurized and no leaks were observed.Microscopic observation revealed no apparent issues.As the no leaks were observed, the complaint is unconfirmed.No further action is required because the reported event could not be related to a deficiency in product manufacture or deficiency while under correct product use.A lot history review (lhr) of reau1602 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (reau1602) have been reported from the same (b)(6) facility.
|
| |
|
Event or Problem Description
|
|
It was reported that patient using picc in the upper left limb showed pain and an increase of the brachial perimeter, edema in the upper left limb.After that the use of the picc was suspended, the case was assessed and measured and reassessed the next day it shows fibrous cord.It was also found that picc has a hole and a solution extravasation."3/6/3018" - it was reported that the "fibrous cord" is a consequence of a superficial thrombosis.It was stated that "the thrombotic process that progressed to fibrosis, which determines the permanence of the hardened venous cord.¿.
|
| |
|