MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751PNAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751PNAL
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pump received with cracked battery tube thread, reservoir tube lip completely broken off, missing reservoir tube lip o-ring, cracked case near display window corners, cracked on display window and minor scratches on display window noted.The test reservoir does not stay locked in place due to reservoir tube lip broken off.
 
Event or Problem Description
Customer reported via phone call that insulin pump had ring lip part of the reservoir compartment broke off.Customer's blood glucose value was 188 mg/dl at the time of the incident.The customer was assisted with troubleshooting.Customer will return device for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751PNAL
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7329350
Report Number3004209178-2018-55945
Device Sequence Number119872
Product Code OZO
UDI-Device Identifier00643169507333
UDI-Public(01)00643169507333
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial
Report Date (Section B) 03/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-751PNAL
Device Catalogue NumberMMT-751PNAL
Device Lot NumberA3751PNALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2018
Initial Report FDA Received Date03/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age54 YR
Patient Weight172
-
-