The associated complaint devices were not returned.The clinical/medical team concluded, all documents and images provided as of this date have been reviewed and considered and unless noted do not contribute to the clinical investigation.Based on the limited information provided, the root cause of this patient¿s pain cannot be concluded, and if the cup position and if any impingement is present along with the reported calcification deposits.However, it could be attributed to the ¿calcific tendinopathy associated with loss of function as result of the reduced offset in the hip after the primary surgery as stated by the surgeon.It has been communicated that no further information would be forthcoming at this time.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed parts revealed no prior complaints for the listed batches.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.
|