SMITH & NEPHEW, INC. TANDEM INTL BIPOLAR 54OD 28ID; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED
|
Back to Search Results |
|
| Catalog Number |
71324054 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Bone Fracture(s) (1870)
|
| Date of Event |
02/07/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Stem subsidence owing to fracture at calcar and greater trochanter surgeon dissected down to hip joint.Dislocated hip.Removed stem and bipolar head as one construct using the polarstem removal tool prepared canal for echelon cemented stem cemented size 14 175mm echelon stem std collar.Trialed bipolar head and hip deemed stable.Inserted definitive bipolar construct onto stem.Reduced hip.Reduced gt fracture and fixated with periprosthetic plate.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected tandem intl cocr shell liner bipolar was not returned for evaluation.A clinical evaluation noted the provided image of an x-ray pelvic overview, dated 3 days before the revision and 2 weeks after implantation, shows the mentioned calcar fracture.As no immediate post-operative x-rays is available for comparison, the subsidence cannot be further described.As the fracture occurred a short time after implantation, an intra-operative peri-prosthetic fracture could have also been unrecognized and simply propagated with progressive weight bearing during the post-operative period.Based on the performed investigation the root cause of the reported of the reported event cannot be determined.A review of the production documentation for the corresponding product did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Without the actual product involved, our investigation of this report is inconclusive.If the device or new information is received in the future, this complaint can be re-opened.
|
| |
|
Search Alerts/Recalls
|
|
|