MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TANDEM INTL BIPOLAR 54OD 28ID; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TANDEM INTL BIPOLAR 54OD 28ID; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED Back to Search Results
Catalog Number 71324054
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 02/07/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Stem subsidence owing to fracture at calcar and greater trochanter surgeon dissected down to hip joint.Dislocated hip.Removed stem and bipolar head as one construct using the polarstem removal tool prepared canal for echelon cemented stem cemented size 14 175mm echelon stem std collar.Trialed bipolar head and hip deemed stable.Inserted definitive bipolar construct onto stem.Reduced hip.Reduced gt fracture and fixated with periprosthetic plate.
 
Additional Manufacturer Narrative
The affected tandem intl cocr shell liner bipolar was not returned for evaluation.A clinical evaluation noted the provided image of an x-ray pelvic overview, dated 3 days before the revision and 2 weeks after implantation, shows the mentioned calcar fracture.As no immediate post-operative x-rays is available for comparison, the subsidence cannot be further described.As the fracture occurred a short time after implantation, an intra-operative peri-prosthetic fracture could have also been unrecognized and simply propagated with progressive weight bearing during the post-operative period.Based on the performed investigation the root cause of the reported of the reported event cannot be determined.A review of the production documentation for the corresponding product did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Without the actual product involved, our investigation of this report is inconclusive.If the device or new information is received in the future, this complaint can be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TANDEM INTL BIPOLAR 54OD 28ID
Common Device Name
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7329703
Report Number1020279-2018-00291
Device Sequence Number16172967
Product Code KWY
UDI-Device Identifier03596010494450
UDI-Public03596010494450
Combination Product (Y/N)N
PMA/510(K) Number
K023743
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date08/08/2026
Device Catalogue Number71324054
Device Lot Number16HM06533
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/08/2018
Supplement Date Received by Manufacturer02/08/2018
Initial Report FDA Received Date03/11/2018
Supplement Report FDA Received Date06/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-