The affected cpcs collarless polished cemented standard offset stem was not returned for evaluation.A clinical evaluation noted based on the limited information provided, the root cause of this patient¿s pain cannot be concluded, and if the cup position and if any impingement is present along with the reported calcification deposits.However, it could be attributed to the ¿calcific tendinopathy associated with loss of function as result of the reduced offset in the hip after the primary surgery¿.It has been communicated that no further information would be forthcoming at this time.A review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Without the actual product involved, our investigation of this report is inconclusive.If the device or new information is received in the future, this complaint can be re-opened.
|