(b)(6).Surgical, inc (isi) received the instrument involved with the complaint and completed the device evaluation.Failure analysis confirmed the customer reported failure mode.The instrument was found to have a broken pitch cable at the distal clevis hub.The broken cable segment that contains the crimp was missing from clevis.The clevis did not exhibit any damage or wear marks.Other cables at the wrist were not damaged.A device history record (dhr) review for the device involved with the complaint has been completed.No non-conformancies were identified to be related to this complaint.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the failure mode were to recur.
|
It was reported that during a da vinci-assisted surgical procedure, the small grasping retractor instrument had a "wire exposed".There was no report of fragments falling inside the patient.The planned surgical procedure was completed and no patient harm, adverse outcome or injury was reported.
|