• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC COLIBRI II HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC COLIBRI II HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.101
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/19/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Device manufacture date: the device manufacture date is unavailable.Therefore, (b)(4).The manufacturing location is currently not available.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was discovered that the small battery drive device did not work.It was not reported if there were any delays to the surgical procedure.A spare device was available to complete the surgery.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The manufacturer location was documented as unknown in the initial report.The location has been updated to waldenburg contact office name/address has been updated accordingly to reflect the corrected manufacturing facility.Device manufacture date: the device manufacture date was unknown in the initial report.The device manufacture date has been updated as 7/2/2013.The udi has been updated accordingly.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Correction: d4: incorrect serial number: the device serial number was incorrect in the initial report.The serial number has been updated from (b)(4).D4: the udi has been updated accordingly.This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.It was determined that the motor seized, jammed and was moving heavily.It was further determined that the device had no function.The assignable root cause was determined to be due to improper maintenance, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COLIBRI II HANDPIECE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7330357
MDR Text Key102119667
Report Number8030965-2018-51918
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.101
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2018
Initial Date Manufacturer Received 02/21/2018
Initial Date FDA Received03/12/2018
Supplement Dates Manufacturer Received06/21/2018
06/27/2018
Supplement Dates FDA Received06/26/2018
07/03/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-