| Model Number |
60729369 |
| Medical Device Problem Code |
Device Issue (2379)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
02/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Bayer quality assurance product analysis received and examined the returned syringe, lot #8538830.Visual examination of the returned product confirmed the presence of a strand of hair-like particle in the syringe barrel.The hair-like material was most likely introduced during the manufacturing process.A 12 month review of complaint history determined the frequency to be (b)(4) complaints per million (cpm) for similar defects.
|
| |
|
Event or Problem Description
|
|
The site reported the following: a stellant sds-ctp-scs syringe was filled with contrast medium using a spike into the contrast medium vial and after filling a hair-like foreign particle was noted to be floating in the contrast inside the syringe.The syringe kit was removed from the area and was not used for a patient.
|
| |
|