• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC CXI SUPPORT CATHETER; KRA CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC CXI SUPPORT CATHETER; KRA CATHETER, CONTINUOUS FLUSH Back to Search Results
Catalog Number CXI-2.6-18-150-P-NS-ANG
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time. a follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that during an unspecified procedure, the cxi support catheter hub fell off during contralateral advancement of the catheter from the left femoral artery to the severely stenosed lesion.Another cxi support catheter of the same lot was utilized to successfully complete the procedure.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Per information received on 14mar2018, the catheter was advanced from the left femoral artery to the common iliac artery (cia) with severe chronic total occlusion (cto) by retrograde contralateral approach.The physician stated that he may have applied too much torque to the device.Pre-existing conditions: cto in the cia.This report includes information known at this time. a follow-up report will be submitted should additional relevant information become available.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation.A review of the dimensional verification, complaint history, device history record, documentation, drawing, quality control, specifications, and a visual inspection and functional evaluation of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed that the proximal fitting (hub) was separated from the shaft of the catheter.The strain relief was removed from the hub for further examination, and it was discovered that a small fragment of the catheter material was retained in the hub.The proximal separated end of the catheter was twisted, which prevented the advancement of a test guide wire.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.Investigation ¿ evaluation.A review of the dimensional verification, complaint history, device history record, documentation, drawing, quality control, specifications, and a visual inspection and functional evaluation of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed that the proximal fitting (hub) was separated from the shaft of the catheter.The strain relief was removed from the hub for further examination, and it was discovered that a small fragment of the catheter material was retained in the hub.The proximal separated end of the catheter was twisted, which prevented the advancement of a test guide wire.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CXI SUPPORT CATHETER
Common Device Name
KRA CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7330927
Report Number1820334-2018-00668
Device Sequence Number13854371
Product Code KRA
UDI-Device Identifier00827002500037
UDI-Public(01)00827002500037(17)201103(10)8345933
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K072724
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCXI-2.6-18-150-P-NS-ANG
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/06/2018
Supplement Date Received by Manufacturer03/14/2018
05/02/2018
Initial Report FDA Received Date03/12/2018
Supplement Report FDA Received Date03/14/2018
05/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-