Catalog Number 490067
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Event 2: this report has been identified as b.Braun medical internal report number (b)(4).One (1) unopened sample of primary set 490067, from lot 61584974, and one (1) used extension set, which was part of a primary set, without packaging was returned for evaluation.The whole sample and the extension set were tested for leakage per specification with passing results.The set was primed with normal saline and no air in line/bubbles were observed in any part of the tubing.In addition the iso luer/taper test was performed on all applicable parts with passing results.Therefore, the reported defect was not confirmed.Review of the discrepancy management system database performed for the reported lot number did not reveal any abnormalities or non-conformances of this nature.We will maintain this report for further references and continue to monitor other reports for similar occurrences.
 
Event or Problem Description
As reported by the user facility: event 2: while the patient was receiving saline kvo, leakage occurred between the extension set and the iv set.The clinician also noticed air bubbles going back into the iv set.The nurse stated that there was a crack in the hub of the iv set which was connected to the extension set, and this crack allowed air to get into the iv set.No patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV ADMINISTRATION SET
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN DOMINICAN REPUBLIC INC.
las americas industrial park
km22 autopista las americas
santo domingo,
DR 
MDR Report Key7330954
Report Number9614279-2018-00059
Device Sequence Number104800
Product Code FPA
UDI-Device Identifier04046964186042
UDI-Public04046964186042
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K083723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 03/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Catalogue Number490067
Device Lot Number0061584974
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2018
Initial Report FDA Received Date03/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-