• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COSMAN MEDICAL CSK ELECTRODE; PROBE, RADIOFREQUENCY LESION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COSMAN MEDICAL CSK ELECTRODE; PROBE, RADIOFREQUENCY LESION Back to Search Results
Model Number CSK-TC15
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/14/2018
Type of Reportable Event Malfunction
Event or Problem Description
Device analysis was performed on the returned electrode found that the shaft was completely separated from the hub due to excessive external mechanical force.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSK ELECTRODE
Common Device Name
PROBE, RADIOFREQUENCY LESION
Manufacturer (Section D)
COSMAN MEDICAL
22 terry ave
burlington MA 01803
MDR Report Key7330977
Report Number3006630150-2018-00984
Device Sequence Number13478660
Product Code GXI
UDI-Device Identifier00813250010114
UDI-Public00813250010114
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K050084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model NumberCSK-TC15
Device Lot Number041017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2018
Initial Report FDA Received Date03/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-