| Model Number |
H7493926020270 |
| Medical Device Problem Codes |
Device Or Device Fragments Location Unknown (2590); Device Damaged by Another Device (2915); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
02/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.(b)(4).Device evaluated by mfr.: the device was not received for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is caused by other device as another device/drug/subsequent procedure caused the complaint event.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that stent dislodgment occurred.The target lesions were located in the left main artery and proximal left anterior descending (lad) artery.A 2.75 x 20 synergy ii drug-eluting stent was advanced to treat the proximal lad lesion but difficulties were encountered in delivering the device.So, the physician treated the left main artery first with a 2.5 x 16 synergy drug-eluting stent.Subsequently, the 2.75 x 20 synergy ii drug-eluting stent was advanced again but the physician still had some difficulties delivering the device.During removal, the physician felt like the stent struts became stuck to the previously implanted stent and after the device was removed from the patient, it was noted that the stent came off the balloon.The stent was nowhere to be found and the physician decided to end the procedure.No further patient complications were reported and the patient's status at the moment was stable.
|
| |
|