• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number H7493926012400
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.(b)(4).
 
Event or Problem Description
It was reported that a foreign matter was present on the device.A 4.00 x 12mm synergy ii stent was selected for use.During preparation, when the physician took the protective sheath off from the device, it was noted that there was something "suspicious" on the stent.The procedure was completed with a different device.No patient complications were reported as the device did not enter the patient's body.
 
Event or Problem Description
It was reported that a foreign matter was present on the device.A 4.00 x 12mm synergy ii stent was selected for use.During preparation, when the physician took the protective sheath off from the device, it was noted that there was something "suspicious" on the stent.The procedure was completed with a different device.No patient complications were reported as the device did not enter the patient's body.
 
Additional Manufacturer Narrative
Device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes updated.Device evaluated by mfr.: stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.A 360-degree review of the stent was carried out both visually and microscopically and no foreign matter was noted on the stent.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed signs of damage.A visual and tactile examination of the hypotube found multiple hypotube kinks along the full length of the catheter.This type of damage is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7331403
Report Number2134265-2018-02077
Device Sequence Number10790748
Product Code NIQ
UDI-Device Identifier08714729840077
UDI-Public08714729840077
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/02/2018
Device Model NumberH7493926012400
Device Catalogue Number39260-1240
Device Lot Number0020611381
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2018
Initial Date Received by Manufacturer 02/14/2018
Supplement Date Received by Manufacturer03/28/2018
Initial Report FDA Received Date03/12/2018
Supplement Report FDA Received Date04/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-