• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH MOBERLY PORTA SAMPLE COLD PK 3X8.25; PACK, HOT OR COLD, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH MOBERLY PORTA SAMPLE COLD PK 3X8.25; PACK, HOT OR COLD, DISPOSABLE Back to Search Results
Model Number 30635-611
Health Effect - Clinical Code Chemical Exposure (2570)
Date of Event 02/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A device history record (dhr) review could not be completed as the lot number was not provided.At the time of this investigation no samples were available, therefore, we could not determine a failure mode or definitive root cause.The plant¿s quality system mandates suitable in-process controls to measure seal integrity of representative samples.Forty-eight samples were pulled during production of each of these lots and tested at predetermined locations to best gauge the seal integrity by pull test and functional leak testing.All samples pulled from these lots met the predetermined criteria before they were released.This product is to be used in the upright position.A corrective action will not be initiated at this time since no samples were provided and the information is not sufficient to identify a root cause.The manufacturing site will continue to monitor the incoming complaints for burst against this product.
 
Event or Problem Description
Based on information received from the customer, on (b)(6) 2018 a phlebotomist sustained a splash to the eyes after a porta sample cold pack ruptured upon activation.The phlebotomist flushed his eyes and reported to employee health.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTA SAMPLE COLD PK 3X8.25
Common Device Name
PACK, HOT OR COLD, DISPOSABLE
Manufacturer (Section D)
CARDINAL HEALTH MOBERLY
808 w highway 24
moberly MO 65270
MDR Report Key7331731
Report Number1423537-2018-00167
Device Sequence Number11170170
Product Code IMD
UDI-Device Identifier50630140015685
UDI-Public50630140015685
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model Number30635-611
Device Catalogue Number30635-611
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2018
Initial Report FDA Received Date03/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age21 YR
Patient Weight91
-
-