Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual examination revealed that the polyaxial¿s tulip head had become dislodged from the polyaxial¿s screw shank.It was also noted that the flex ball had become partially damaged and deformed.Swage markings were present indicative of the cap being properly placed in its intended position within the tulip head.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.A definitive root cause for the polyaxial¿s screw tulip head becoming dislodged cannot be positively determined.However, visual evaluation suggests that unanticipated high amount of forces was placed on the tulip head, resulting in the tulip head becoming dislodged from the shaft.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|