SMITHS MEDICAL ASD, INC. JELCO® PROTECTIV® SAFETY I.V. CATHETER; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
|
Back to Search Results |
|
| Catalog Number |
305706 |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
Discomfort (2330)
|
| Date of Event |
02/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a jelco® protectiv® safety i.V.Catheter was dull, causing patient "discomfort".No injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One used locked needle guard assembly was received for investigation.The actually catheter was not returned for evaluation and therefore could not be examined or evaluated.The returned sample was examined under magnification and photographed.The photos display evidence of dried blood residue in cannula point and flash chamber.There was no visual evidence of any burrs, hooks or any potential needle point damaged.As part of the investigation, the device was manually unlocked, the sample advanced smoothly, clicked, and locked as designed with no manufacturing nonconformances.
|
| |
|
Search Alerts/Recalls
|
|
|