| Model Number |
111981 |
| Medical Device Problem Codes |
Difficult to Open or Close (2921); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The device displayed nicks on the knife blade.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Replication of the observed knife blade damage may occur if the instrument is applied over an obstruction.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.The information booklet cautions the user to make certain the section of the tissue to be stapled is free from any metal clips or similar structures; otherwise, the knife blade may not cut.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during an open procedure of low anterior resection, the device was difficult or unable to close.Unusual effort was needed to turn the twist knob in order to visualize the green bar.The patient had no injury.
|
| |
|