• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COSMAN MEDICAL RF CANNULA; PROBE, RADIOFREQUENCY LESION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COSMAN MEDICAL RF CANNULA; PROBE, RADIOFREQUENCY LESION Back to Search Results
Model Number CC101021
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hypersensitivity/Allergic reaction (1907)
Date of Event 01/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was discarded and will not be returned.It is indicated that the device will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.A review of the manufacturing documentation for the device revealed that no anomalies or deviations potentially related to the event occurred during manufacturing.A review of the sterilization documentation for the device was found to be satisfactory.
 
Event or Problem Description
A report was received that one week after a radiofrequency procedure the patient experienced a reaction that was initially assessed to be burn marks at the puncture site.The size of the marks were less than 1 centimeter and were located at l3-s1.During the procedure, it was noted that the physician injected 1-2 ml of trigon 400 mg/ml with saline through the cannulas.They did not use any skin preparation or dressing that could have gotten into the puncture site.The physician felt that the marks were related to use of the cannula, and later assessed that he doesn¿t know the cause of the reaction but believes it could be an allergic reaction.The patient was administered a two month prescription for corticoid ointment the patient is doing fine.There is no further course of action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RF CANNULA
Common Device Name
PROBE, RADIOFREQUENCY LESION
Manufacturer (Section D)
COSMAN MEDICAL
22 terry ave
burlington MA 01803
MDR Report Key7333024
Report Number3006630150-2018-00989
Device Sequence Number144648
Product Code GXI
UDI-Device Identifier00813250012484
UDI-Public00813250012484
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K060799
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date06/08/2021
Device Model NumberCC101021
Device Lot NumberL098
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2018
Initial Report FDA Received Date03/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-