| Model Number |
1007CC |
| Health Effect - Clinical Code |
Heart Failure (2206)
|
| Date of Event |
10/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device arrived in good condition.There were no signs of any damage to the device.The customer complaint that the device would not shut off could not be confirmed.Upon the initial evaluation the compressor could be heard turning on and off when the switch was turned.When the depth compression knob the device began to compress, the on/off switch was turned to off and the compression stopped.The front cover was taken off the device to examine the on/off switch but there is no damage to the assembly.Long term testing was not able to reproduce the complaint.Unit was returned to the customer.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was used to perform cpr on a patient for about 30 minutes.The patient was not revived but when attempting to remove the device, it would not shut off.The oxygen hose was disconnected which stopped the unit.It was stated that the problem did not contribute to the failure to revive the patient.
|
| |
|