A field service engineer (fse) was dispatched, fse ordered arrived at the site.The customer's information technology (it) department stated that they would assist her with the installation the probable cause of the reported event was due to clogged i-6 filter housing peek tubing.The customer also reported that the pc was finally up and running.Fse assisted her with the set up; however, the customer reported that she was still receiving errors fse followed up with the customer again to verify that she was up and running.There were no further reports of errors.No further action was required by field service.The probable cause of the reported event was due to fault/ failure of the external cpu.(b)(4).This report is being submitted due to a retrospective review conducted under capa (b)(4).
|
On (b)(6) 2016, the customer reported receiving a "host comm error" mid-run, with their g8 analyzer.A field service engineer (fse) was dispatched on (b)(6) 2016, to address the reported event,which resulted in delay in reporting of patient results hba1c.There was no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|