• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. FOUNDATION HIP; SHELL, ACETABULAR, FLARED FMP, NO HOLES SIZE52

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. FOUNDATION HIP; SHELL, ACETABULAR, FLARED FMP, NO HOLES SIZE52 Back to Search Results
Model Number 430-04-052
Health Effect - Clinical Code Pain (1994)
Date of Event 03/07/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to pain.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to the pain.The previous surgery and the revision detailed in this investigation occurred over 9 years and 9 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to the pain.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are many factors that may contribute to the event that are outside the control of djo surgical such as: excessive loading, degenerative bone disease, patient bone deterioration, soft tissue impingement, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION HIP
Common Device Name
SHELL, ACETABULAR, FLARED FMP, NO HOLES SIZE52
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7334089
Report Number1644408-2018-00217
Device Sequence Number646032
Product Code LPH
UDI-Device Identifier00888912082389
UDI-Public(01)00888912082389
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K974095
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2012
Device Model Number430-04-052
Device Catalogue Number430-04-052
Device Lot Number53837562
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/07/2018
Supplement Date Received by Manufacturer04/12/2018
Initial Report FDA Received Date03/13/2018
Supplement Report FDA Received Date04/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
497-34-000, LOT 53889360; 499-34-007, LOT 53777187; 497-34-000, LOT 53889360; 499-34-007, LOT 53777187
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age85 YR
-
-