Model Number H7493952838300
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.  (b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.A 3.00x38mm promus premier¿ stent was selected for use.The device was loaded onto the wire and while inserting it into the sheath, it was noted that the shaft broke.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes updated.Device evaluated by mfr: the stent delivery system (sds) was returned for analysis.A visual examination of the stent found no damages.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The crimped stent outer diameter (od) was measured and is within maximum crimped stent profile measurement.A visual and microscopic examination of the bumper tip found no signs of damage.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the device found a break in the hypotube 637 mm distal from the distal end of the strain relief.Multiple hypotube kinks were also noted along the full length of the catheter.This type of damage is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the shaft polymer extrusion found a kink in the midshaft 25 mm distal to the proximal end of the hypotube midshaft bond.No other issues were identified during the product analysis.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.A 3.00x38mm promus premier¿ stent was selected for use.The device was loaded onto the wire and while inserting it into the sheath, it was noted that the shaft broke.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER¿
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7334144
Report Number2134265-2018-01733
Device Sequence Number11075608
Product Code NIQ
UDI-Device Identifier08714729845027
UDI-Public08714729845027
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P110010
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/07/2019
Device Model NumberH7493952838300
Device Catalogue Number39528-3830
Device Lot Number21392697
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2018
Initial Date Received by Manufacturer 02/16/2018
Supplement Date Received by Manufacturer03/28/2018
Initial Report FDA Received Date03/13/2018
Supplement Report FDA Received Date04/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-