• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. CLP HIP; HIP, STEM, CLP, STANDARD, SIZE#3, IMPLANT, POLISHED NECK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. CLP HIP; HIP, STEM, CLP, STANDARD, SIZE#3, IMPLANT, POLISHED NECK Back to Search Results
Model Number 426-10-030
Health Effect - Clinical Code No Code Available (3191)
Date of Event 03/03/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient having instability.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to instability.The previous surgery and the revision detailed in this investigation occurred 4.5 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to this event.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.The root cause of this complaint was a revision surgery due to instability.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLP HIP
Common Device Name
HIP, STEM, CLP, STANDARD, SIZE#3, IMPLANT, POLISHED NECK
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7334623
Report Number1644408-2018-00208
Device Sequence Number10764435
Product Code JDI
UDI-Device Identifier00888912079518
UDI-Public(01)00888912079518
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K932481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/28/2021
Device Model Number426-10-030
Device Catalogue Number426-10-030
Device Lot Number403N1030
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/03/2018
Supplement Date Received by Manufacturer04/06/2018
Initial Report FDA Received Date03/13/2018
Supplement Report FDA Received Date04/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-323, LOT 862B1104; 400-03-323, LOT 862B1104
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
-
-