Model Number EM210RF
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation was not performed; the device was not returned for analysis.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Customer reported the handpiece was getting warm.There was no patient impact.
 
Additional Manufacturer Narrative
Summary of analysis: the reported malfunction could not be replicated.The evaluation confirmed there was no fault found with the device.The ambient temperature was 74.9f and the rise or max temperature was 89.7f.The max temperature - the ambient temperature = 14.8f, which was passing.The max rise temperature was 27f for the drill.The item was repaired cleaned, tested and passed all manufacturing specifications.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPC® HANDPIECE - STYLUS TOUCH®
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7334676
Report Number1045254-2018-00107
Device Sequence Number41042
Product Code HBE
UDI-Device Identifier00613994790217
UDI-Public00613994790217
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K081475
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEM210RF
Device Catalogue NumberEM210RF
Device Lot Number214053382
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2018
Supplement Date Received by Manufacturer03/21/2018
Initial Report FDA Received Date03/13/2018
Supplement Report FDA Received Date03/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No