• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS, INC SOFT-VU SIZING ANGIOGRAPHIC CATHETER; SOFT-VU SIZING ANGIOGRAPHIC CATHETER,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS, INC SOFT-VU SIZING ANGIOGRAPHIC CATHETER; SOFT-VU SIZING ANGIOGRAPHIC CATHETER, Back to Search Results
Catalog Number H787107120015
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 02/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported defective device has yet to be returned to the manufacturer for a device evaluation.The firm is attempting to obtain the device.An investigation into the root cause for product problem is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.(b)(4).
 
Event or Problem Description
As reported to angiodynamics on (b)(6) 2018: it was reported that during an angiogram visceral artery procedure, the angiographic catheter fractured inside of the patient.The catheter was removed, and the fragment was successfully retrieved from inside of the patient.The procedure was completed using another device.It was reported the patient suffered no permanent harm or injury due to this event.It was indicated the defective disposable device is available for return to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
As the reported defective device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of catheter fracture cannot be confirmed.As a device evaluation was not performed, a root cause for the reported complaint description cannot be determined.Returned was a soft vu catheter.A visual examination of the returned device noted the catheter tip was separated from the catheter shaft.The device met all dimensional specifications for this product.There was no evidence of of material stretching.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use, which is supplied to the end user with this device, states; "the maximum pressure limit of catheters intended for flush angiography is stated on the catheter package.When using a pressure injector, do not exceed the stated maximum pressure." the ifu also states; "do not attempt to hand straighten the tip of any curved tipped catheters which are furnished with a tip straightener.This may result in damage to the product.Tip straighteners are furnished on all catheters intended to be straightened with the aid of a tip straightener.Reshaping of the catheter tip is not recommended.Physical damage to the catheter material can result when exposed to heat." a review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFT-VU SIZING ANGIOGRAPHIC CATHETER
Common Device Name
SOFT-VU SIZING ANGIOGRAPHIC CATHETER,
Manufacturer (Section D)
ANGIODYNAMICS, INC
603 queensbury avenue
queensbury NY 12804
MDR Report Key7334811
Report Number1319211-2018-00027
Device Sequence Number51705
Product Code DQO
UDI-Device IdentifierH787107120015
UDI-PublicH787107120015
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K112452
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/31/2018
Device Catalogue NumberH787107120015
Device Lot Number4919082
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2018
Supplement Date Received by Manufacturer03/05/2018
Initial Report FDA Received Date03/13/2018
Supplement Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-