Returned for evaluation was one guidewire.A visual review of the returned device noted what appeared to be biological matter.The guidewire coil appeared to be stretched.The supplier of this device was notified of the event via (b)(4).The sample was forwarded to the supplier for a device evaluation and root cause determination.The customer's reported complaint description of the guidewire unraveling is confirmed.Per the supplier's response, "the physical analysis is unable to determine the cause of failure, however foreign mass is likely a contributing factor of the coil stretching".The supplier's review of the device's manufacturing records noted no deviations or observations at the time of manufacture.Although it is confirmed the supplier states "the physical analysis is unable to determine the cause of failure; however foreign mass is likely a contributing factor of the coil stretching".A review of the lot history records was performed for packaging lot (pl01417) for catalog number h787058001135 for any deviations related to the reported defect of the complaint.The review confirmed that the packaging lots and all component lots met all material, assembly, and performance specification.The instructions for use, which is supplied to the end user with this catalog number (sp-3363) inspect the guidewire prior to use for tip shape, bends, kinks or coil separation.Do not use if damaged.Using sterile technique, localize and puncture the vessel with a needle cannula.Remove the needle, leaving the cannula in place.Insert distal end of the guidewire through the cannula and into the vessel.Remove cannula, leaving guidewire within the lumen of the vessel.Pass the catheter over the guidewire within the lumen of the vessel under fluoroscope guidance to the desired position.Carefully remove the guidewire from the catheter.Excessive force against resistance may result in damage to guidewire and catheter or vessel perforation.Angiodynamics does not recommend a particular technique for the use of this device.The steps contained in the preceding directions discuss the seldinger technique for percutaneous entry and are for information purposes only.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint reference (b)(4).
|