Catalog Number 0006428
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The full number for the product in question is (b)(4) (1/27/03).Immucor technical support used a remote electronic connection method on (b)(6) 2018 to assess the instrument test well image in question, and determined that the instrument test well image in question was visually negative.An immucor field service engineer visited the customer site to assess the testing instrument used on 28feb2018 and found it to be operating as expected.
 
Event or Problem Description
On (b)(6) 2018, a (b)(6) customer site reported an unexpected negative antibody screen when using capture-r ready indicator red cells on a galileo neo instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE-R READY INDICATOR RED CELLS
Common Device Name
ANTIGLOBULIN COATED RED CELLS
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 5625
MDR Report Key7335507
Report Number1034569-2018-00062
Device Sequence Number74328
Product Code KSF
UDI-Device Identifier10888234000600
UDI-Public10888234000600
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
BK020053
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Medical Technologist
Type of Report Initial
Report Date (Section B) 03/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Medical Technologist
Device Expiration Date02/19/2018
Device Catalogue Number0006428
Device Lot Number221019
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2018
Initial Report FDA Received Date03/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
-
-