| Catalog Number |
0006428 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The full number for the product in question is (b)(4) (1/27/03).Immucor technical support used a remote electronic connection method on (b)(6) 2018 to assess the instrument test well image in question, and determined that the instrument test well image in question was visually negative.An immucor field service engineer visited the customer site to assess the testing instrument used on 28feb2018 and found it to be operating as expected.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a (b)(6) customer site reported an unexpected negative antibody screen when using capture-r ready indicator red cells on a galileo neo instrument.
|
| |
|