Additional narrative: there was no known reported patient involvement associated with the complained event.Unknown when device malfunctioned.(b)(4).Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.(b)(4).The investigation could not be completed; no conclusion could be drawn, as no product was received.A review of the device history records has been requested and is currently pending completion.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Date device received at manufacturing site for investigation.A device history record (dhr) review was performed for part no.: 03.617.904, lot no.: l392053 , manufacturing location: (b)(4), release to warehouse date: 20.Jul.2017: no non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.A product investigation was performed.The screw inserter (part 03.617.904, lot l392053, mfg 20.Jul.2017) was received at us cq with a stripped distal tip.The complaint condition was visually confirmed and consistent with the reported condition.Replication of the complaint condition is not applicable as the devices were returned damaged.The complaint condition is confirmed as screw inserter (part 03.617.904, lot l392053) was received stripped.A device inspection, device history record review and drawing review were performed as part of this investigation.There is no indication that a design or manufacturing issue contributed to the complaint.While no definitive root cause could be determined it is possible that the device encountered unintended forces (such as being over torqued).No new malfunctions were observed during the course of this investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|