| Model Number |
ZM-541PA |
| Medical Device Problem Codes |
Overheating of Device (1437); Human Factors Issue (2948); Temperature Problem (3022); Misassembly by Users (3133)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer reported that the transmitter overheated.The device was sent in for evaluation and they were provided with an exchange.The batteries required for this investigation were not returned.The battery cover was also missing.When new batteries were inserted into the unit, heating could not be duplicated.Inspection of the spring contacts shows one badly bent spring and the inside shield showed signs of fluid intrusion.Both of these observations could contribute to the overheating observed (i.E.Fluid with improper battery insertion).
|
| |
|
Event or Problem Description
|
|
The customer reported that the transmitter overheated.
|
| |
|
Additional Manufacturer Narrative
|
|
Details of the complaint: on 02/12/18, customer at community health network reported the transmitter (zm-541pa sn: (b)(6) was getting hot.Service requested/performed exchange.Investigation result: per nkc dhr, the unit has had no history of ncmr, deviation, or capa during manufacturing of the device.The device has not been refurbished and there were no discrepancies or unusual findings before device release that might relate to the reported issue.The device warranty began on 04/23/15, and the issue occurred after almost 3 years at the customer facility.Review of device c4c history found no additional issue reported for the unit.Evaluation of the unit at nka was unable to reproduce the heating.The spring contacts were found to have one badly bent spring, and the inside shield showed signs of fluid intrusion.Zm-540pa operator's manual advises the following: this transmitter is not waterproof.If detergent or liquid spills into the transmitter, stop using it and contact your nihon kohden representative.If a wet transmitter is used, the patient or operator may receive an electrical shock or injury.The root cause is determined to be damage to the unit due to fluid intrusion.Investigation conclusion: the root cause is determined to be damage to the unit due to fluid intrusion.
|
| |
|
Event or Problem Description
|
|
The customer reported that the transmitter overheated.
|
| |
|