• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE FOLEY CATHETER Back to Search Results
Model Number 166816
Medical Device Problem Code Mushroomed (2987)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Information (3190)
Date of Event 01/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the user could not keep the balloon inflated during a period of two days.The balloon lost 30cc in that time.Per the sample evaluation, there was a mushroom shaped balloon.
 
Event or Problem Description
It was reported that the user could not keep the balloon inflated during a period of two days.The balloon lost 30cc in that time.Per the sample evaluation, there was a mushroom shaped balloon.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Correction: catalog #.
 
Event or Problem Description
It was reported that the user could not keep the balloon inflated during a period of two days.The balloon lost 30cc in that time.Per the sample evaluation, there was a mushroom shaped balloon.
 
Additional Manufacturer Narrative
The reported event was confirmed as cause unknown.1 used silicone catheter was returned for evaluation.Per evaluation of the sample, a mushroom shaped balloon was confirmed.During the visual inspection a cuff roll was observed.During the functional evaluation a cuff roll formed.The dimensional evaluation the catheter was found within specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "to deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Visually inspect the product for any imperfections or surface deterioration prior to use.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE FOLEY CATHETER
Common Device Name
SILICONE FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7336170
Report Number1018233-2018-00796
Device Sequence Number19200090
Product Code GBM
UDI-Device Identifier00801741029806
UDI-Public(01)00801741029806
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K760093
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Model Number166816
Device Catalogue Number166818
Device Lot NumberNGAY3219
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2018
Initial Date Received by Manufacturer 02/22/2018
Supplement Date Received by Manufacturer03/22/2018
03/28/2018
Initial Report FDA Received Date03/13/2018
Supplement Report FDA Received Date03/23/2018
04/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-