• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD BRACHYTHERAPY, INC. -1424526 QUICKLINK¿ LOADER; QUICKLINK LOADER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD BRACHYTHERAPY, INC. -1424526 QUICKLINK¿ LOADER; QUICKLINK LOADER Back to Search Results
Catalog Number 70310QCA1
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during a 3/4 curietherapie procedure there was a problem with the quicklink applicator.It had blocked due to the failure to tighten a screw located under the rack, preventing the advance of the needle.The screw was loosened during the procedure since the tests and the preparation of 12 needles before had been carried out without difficulty.The customer had to take a new applicator to complete the procedure.There was an extended operating time for the patient of about 10 minutes.
 
Additional Manufacturer Narrative
Upon further review, bard medical has determined that this mdr was initially reported in error as this event is not reportable.
 
Event or Problem Description
It was reported that during a 3/4 curietherapie procedure there was a problem with the quicklink applicator.It had blocked due to the failure to tighten a screw located under the rack, preventing the advance of the needle.The screw was loosened during the procedure since the tests and the preparation of 12 needles before had been carried out without difficulty.The customer had to take a new applicator to complete the procedure.There was an extended operating time for the patient of about 10 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKLINK¿ LOADER
Common Device Name
QUICKLINK LOADER
Manufacturer (Section D)
BARD BRACHYTHERAPY, INC. -1424526
295 east lies road
carol stream IL 60188
MDR Report Key7336173
Report Number1018233-2018-00797
Device Sequence Number1589385
Product Code IWJ
UDI-Device Identifier00801741079238
UDI-Public(01)00801741079238
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number70310QCA1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2018
Initial Date Received by Manufacturer 02/23/2018
Supplement Date Received by Manufacturer03/20/2018
Initial Report FDA Received Date03/13/2018
Supplement Report FDA Received Date03/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
-
-