Model Number MICL12.6
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation: lens was returned in liquid, in lens vial.Visual inspection found a piece of optic and haptic missing.Work order search: no similar complaint type event was reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated that 12.6mm micl12.6,implantable collamer lens, -8.0 diopter, would not load properly/loading issue.There was no patient contact with the product and the back up lens was used with no problem.The reporter stated that the cause of event was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7336334
Report Number2023826-2018-00394
Device Sequence Number1504027
Product Code MTA
UDI-Device Identifier00841542102988
UDI-Public00841542102988
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/30/2019
Device Model NumberMICL12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2018
Initial Report FDA Received Date03/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-