SMITH & NEPHEW, INC. HEALICOIL RG DILATOR 5.5MM DISPOSABLE; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Model Number |
72203952 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during arcr procedure, when surgeon use the dilator on humerus, the handle has came free from shaft and got inside the patient.This shaft was removed using a nipper.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One disposable 5.5 healicoil rg dilator returned.Complaint states: ¿it was reported that during arcr procedure, when surgeon use the dilator on humerus, the handle has came free from shaft and got inside the patient.This shaft was removed using a nipper.¿ visual inspection shows that the shaft has been replaced back into the handle.There are ¿nipper¿ marks where the shaft was removed from the patient and has been forced back into the handle.The patient was reported to have soft bone.Excessive force can result in instrument failure.No root cause related to the manufacturing of this device was established.
|
| |
|