| Model Number |
7211010S |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device was making a loud noise.There was no back up device available and no patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of loud noise could not be reproduced.Product passed functional testing per factory test with no faults or errors.Product passed functional testing during 6 hour burn-in on wet station utilizing low and high pressure and flow settings.Raw and zero transducer readings were normal and well within specs during all functional tests.At no time during functional testing did pump become noisy.Noise level of pump during operation was below normal.All functions perform as expected.Pressure readings were normal throughout burn-in on wet station.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.
|
| |
|