MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REPL CONTROL UNIT DYONICS 25; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REPL CONTROL UNIT DYONICS 25; ARTHROSCOPE Back to Search Results
Model Number 7211010S
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/07/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was making a loud noise.There was no back up device available and no patient injury was reported.
 
Additional Manufacturer Narrative
The device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of loud noise could not be reproduced.Product passed functional testing per factory test with no faults or errors.Product passed functional testing during 6 hour burn-in on wet station utilizing low and high pressure and flow settings.Raw and zero transducer readings were normal and well within specs during all functional tests.At no time during functional testing did pump become noisy.Noise level of pump during operation was below normal.All functions perform as expected.Pressure readings were normal throughout burn-in on wet station.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL CONTROL UNIT DYONICS 25
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key7336709
Report Number1643264-2018-00217
Device Sequence Number11094209
Product Code HRX
UDI-Device Identifier03596010555960
UDI-Public(01)03596010555960
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K171794
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7211010S
Device Catalogue Number7211010S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2018
Initial Date Received by Manufacturer 03/07/2018
Supplement Date Received by Manufacturer04/23/2018
Initial Report FDA Received Date03/13/2018
Supplement Report FDA Received Date04/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-