• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS VERTEX RECONSTRUCTION SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS VERTEX RECONSTRUCTION SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Catalog Number 7753500
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Pre-operative diagnosis: c4-c5 dislocation type of procedure: posterior cervical fixation using verte levels implanted: c2-t2 it was reported that on (b)(6) 2018, intra-op, while final tightening of locking screw using the straight hex torque driver with t-handle the set screw broken into two pieces.The broken set screw was removed and exchanged with the new one, it added 5 minutes extra time to the procedure.No patient complications were occurred as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERTEX RECONSTRUCTION SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key7337841
Report Number1030489-2018-00388
Device Sequence Number18866415
Product Code KWP
UDI-Device Identifier00613994448743
UDI-Public00613994448743
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K082728
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number7753500
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2018
Initial Report FDA Received Date03/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No