Model Number PM3262
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Patient Involvement (2645)
Event Date 03/02/2018
Event Type  malfunction  
Manufacturer Narrative
No conclusion code available; failure event observed during analysis.Final analysis found that when tested at 40°c or higher temperatures, the connection between the feed through ribbon and left ventricular setscrew block becomes intermittent resulting from inadequate weld joint.
 
Event Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7338351
MDR Text Key102443692
Report Number2017865-2018-03670
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial
Report Date 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/31/2019
Device Model NumberPM3262
Device Lot NumberA000043239
Other Device ID Number05414734509091
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2017
Date Manufacturer Received03/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-