MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY STRATA II VALVE, SMALL, WITH BIOGLIDE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY STRATA II VALVE, SMALL, WITH BIOGLIDE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 92856
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code Therapeutic Effects, Unexpected (2099)
Date of Event 03/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that after the procedure was completed, no change in intracranial pressure was observed.The device was replaced by another device.
 
Additional Manufacturer Narrative
Implant, explant, and event dates updated.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
The returned valve was patent.The valve met the requirements for siphon, pressure-flow, and pre-implantation testing.The valve did not meet the requirements for reflux testing.There was proteinaceous debris observed within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open resulting in fluid reflux.Instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ the returned valve also did not meet the requirements for leak due to a tear in the bottom of the delta chamber.It is unknown how or when this damage occurred.The ifu caution that ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting, or crushing of components.¿ the ifu also caution that ¿care should be taken in handling the valves as silicone has a low cut and tear resistance.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA II VALVE, SMALL, WITH BIOGLIDE
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7338927
Report Number2021898-2018-00140
Device Sequence Number1090449
Product Code JXG
UDI-Device Identifier00643169010420
UDI-Public00643169010420
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K042465
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2019
Device Model Number92856
Device Catalogue Number92856
Device Lot NumberE13242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2018
Supplement Date Received by Manufacturer03/15/2018
04/06/2018
Initial Report FDA Received Date03/14/2018
Supplement Report FDA Received Date03/22/2018
04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-