MEDTRONIC NEUROSURGERY PERITONEAL CATHETER WITH BIOGLIDE AND 8 SLITS; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 27536
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code Therapeutic Effects, Unexpected (2099)
Date of Event 03/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that after the procedure was completed, no change in intracranial pressure was observed.The device was replaced by another device.
 
Additional Manufacturer Narrative
Implant, explant, and event dates updated.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
The peritoneal catheter was returned and was patent.However, the returned catheter did not meet the requirements for leak testing because it was returned in two pieces, one measuring 65 cm, and one measuring 18 cm, a total of 83 cm.It is unknown how or when this damage occurred.The edges around the tear were jagged.The instructions for use (ifu) that accompany the device caution that ¿low tear strength is a characteristic of most unreinforced silicone elastomer materials.Care must be taken with the handling and placement of the silicone elastomer catheter tubing to avoid cuts, nicks or tears.¿ proteinaceous debris was observed within the interior and exterior of the catheter.All catheters are 100% inspected at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERITONEAL CATHETER WITH BIOGLIDE AND 8 SLITS
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7338965
Report Number2021898-2018-00141
Device Sequence Number1228244
Product Code JXG
UDI-Device Identifier00643169472808
UDI-Public00643169472808
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K951258
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Model Number27536
Device Catalogue Number27536
Device Lot NumberE27296
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2018
Supplement Date Received by Manufacturer03/15/2018
04/06/2018
Initial Report FDA Received Date03/14/2018
Supplement Report FDA Received Date03/22/2018
04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-