• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS FEM COMPONENT, CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS FEM COMPONENT, CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5515-F-301
Medical Device Problem Codes Device Contamination with Chemical or Other Material (2944); Packaging Problem (3007)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
It was reported that during a primary left tka, outer blister of the femoral component was opened and what appeared to be a hair was on the tyvek of the inner blister.Due to sterility concerns, surgeon elected not to use the implant before it entered the field.Another implant was immediately available and rep reported that no surgical delay was caused.Rep was present at the time of discovery and witnessed the implant opened.
 
Additional Manufacturer Narrative
Conclusions: the investigation concluded the returned packaging to be nonconforming.Visual inspection of the returned product confirmed a small piece of hair on the top of the inner tyvek seal near the edge of the outer blister.Nc was raised to investigate the confirmed issue.
 
Event or Problem Description
It was reported that during a primary left tka, outer blister of the femoral component was opened and what appeared to be a hair was on the tyvek of the inner blister.Due to sterility concerns, surgeon elected not to use the implant before it entered the field.Another implant was immediately available and rep reported that no surgical delay was caused.Rep was present at the time of discovery and witnessed the implant opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS FEM COMPONENT, CEMENTED
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7338969
Report Number0002249697-2018-00744
Device Sequence Number852186
Product Code MBH
UDI-Device Identifier07613327050356
UDI-Public07613327050356
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Catalogue Number5515-F-301
Device Lot NumberBAA3TBAH4D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2018
Initial Date Received by Manufacturer 02/15/2018
Supplement Date Received by Manufacturer06/19/2018
Initial Report FDA Received Date03/14/2018
Supplement Report FDA Received Date07/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age71 YR
Patient Weight82
-
-