• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC MICROPUNCTURE TRANSITIONLESS ACCESS SET; DYB INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC MICROPUNCTURE TRANSITIONLESS ACCESS SET; DYB INTRODUCER, CATHETER Back to Search Results
Catalog Number MPIS-501-NT-SST
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time. a follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that during right femoral access, the micropuncture transitionless access set wire guide was inserted into the set needle and resistance was felt during advancement through the needle.The wire guide then began to separate upon removal through the needle.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.  according to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Additional information regarding event details, patient anatomy and outcome has been requested, but is not available at this time.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the dimensional verification, complaint history, device history record, documentation, manufacturing instructions, quality control, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed that the wire began to unravel distal to the weld.The unraveled wire had a total length of 66.7 cm.The analysis did not provide any evidence that the wire separated.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROPUNCTURE TRANSITIONLESS ACCESS SET
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7339301
Report Number1820334-2018-00781
Device Sequence Number308665
Product Code DYB
UDI-Device Identifier00827002438705
UDI-Public(01)00827002438705(17)201110(10)8368837
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMPIS-501-NT-SST
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 02/20/2018
Supplement Date Received by Manufacturer04/09/2018
Initial Report FDA Received Date03/14/2018
Supplement Report FDA Received Date05/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-