• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONMED CORPORATION OPEN TIP YANKAUER WITH VENT; SURGICAL INSTRUMENTS, STERILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONMED CORPORATION OPEN TIP YANKAUER WITH VENT; SURGICAL INSTRUMENTS, STERILE Back to Search Results
Model Number 0034980
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported unused device was returned for evaluation in its original unopened package.Label verification and visual inspection were performed.A tear was found in the poly material, adjacent to the device's proximal end.This defect resulted in a sterile breach and was likely caused by the device after the sealing process.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of this product.The products released for distribution were found to have met all specifications prior to shipment.Of the lot containing (b)(4) units, one other complaint has been received.(b)(4).This reported packaging issue was obvious to the distributor, prompting the return of the device for evaluation.There was no patient involvement.As with all medical devices, examination of the product occurs multiple times prior to use.Good clinical practice would include examination and verification of the original packaging and its labeling to ensure both are intact.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event or Problem Description
The distributor in (b)(4) reported a "tear" on the sterile pouch of device.There was no patient involvement as this defect was discovered during incoming inspection.This report is raised on the basis of a sterility breach found during the device evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN TIP YANKAUER WITH VENT
Common Device Name
SURGICAL INSTRUMENTS, STERILE
Manufacturer (Section D)
CONMED CORPORATION
525 french road
utica NY 13502
MDR Report Key7339394
Report Number1320894-2018-00071
Device Sequence Number592091
Product Code GCX
UDI-Device Identifier10653405000341
UDI-Public(01)10653405000341(17)220122(10)201701231
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date01/22/2022
Device Model Number0034980
Device Catalogue Number0034980
Device Lot Number201701231
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2018
Initial Report FDA Received Date03/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-