• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/02/2016
Event Type  malfunction  
Manufacturer Narrative
On 03-aug-2016, an fse conducted on site follow-up with the customer and flushed the fluidic system, primed and adjusted the flow rate.The low total area persisted, so a periodic maintenance was performed.The most probable cause of the reported event was due to a need to adjust the flow rate.(b)(4).This report is being submitted due to a retrospective review conducted under capa-(b)(4).
 
Event Description
On (b)(6) 2016, a customer reported low total area on the first sample when run diluted and as whole blood on their g8 analyzer.The customer is unable to report hba1c patient samples in this position.A field service engineer (fse) was dispatched to address the reported event.The reported issue with their g8 instrument has the potential of a delay in reporting patient results for hba1c assays.There is no indication of any patient intervention or adverse health consequences due to the potential delay of reporting patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7339458
MDR Text Key102589963
Report Number8031673-2018-03010
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/14/2018
Distributor Facility Aware Date08/02/2016
Device Age2 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/14/2018
Initial Date Manufacturer Received 08/02/2016
Initial Date FDA Received03/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
-
-