| Model Number |
989605406772 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
|
| |
|
Event or Problem Description
|
|
A customer reported that while performing a procedure with an s7-3t model transducer, a portion of the probe¿s tip cracked and the bending neck sheath began to detach from the device.The transducer was promptly removed from the patient¿s throat with no part of the probe remaining in the patient.The site¿s philips field service engineer confirmed there were no injuries associated with this event.The suspect transducer has been removed from service and will be returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the transducer confirmed the condition as described by the customer.The investigation noted damage to the probe tip and control handle, a swollen sheath, fluid ingress in the connector, blue ink on the strain relief, cracked beading, and dead elements with the array.The failure analysis concluded the extensive physical damage to the device was caused by improper chemical exposure from the site.
|
| |
|