BOSTON SCIENTIFIC - SPENCER OPTIFLEX¿; DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number M0063903010
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A visual analysis of the returned device found the basket was in an opened/extended position when received.A functional analysis was performed and found that the thumb wheel would move freely in both directions; however, the sheath would not move over the wire and the basket would not close.Further analysis upon disassembly revealed that the pull wire had been broken at its proximal end.The evaluation concluded that the failure (pull wire broken) could have been caused due to the excessive force, or twisting of the device by the user.This could have affected the device's ability to retract the basket.Based on the information available it is possible that the way in which the device was handled and manipulated during procedure may have contributed to the failures noted, therefore the most probable cause of this complaint is operational context, probably due to anatomical/procedural factors encountered during the procedure, performance was limited.The device history record (dhr) review found the device met all manufacturing specifications.A search of the complaint database revealed that no similar complaints exist for the specified lot.
 
Event or Problem Description
It was reported to boston scientific corporation that an optiflex¿ basket was used in a ureteroscopy (urs) stone removal procedure to help treat calyceal calculi located in the upper ureter, pelvic and calyx, on (b)(6) 2018.According to the complaint, during the procedure, the optiflex basket was utilized to retrieve a 2 cm stone.However, the basket could not open or close freely.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be "stable." this event has been deemed a reportable event based on the investigation results; pull wire broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLEX¿
Common Device Name
DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
MDR Report Key7339793
Report Number3005099803-2018-00611
Device Sequence Number19118617
Product Code FFL
UDI-Device Identifier08714729772859
UDI-Public08714729772859
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2018
Device Model NumberM0063903010
Device Catalogue Number390-301
Device Lot Number19514551
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2018
Initial Report FDA Received Date03/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age45 YR
Patient Weight62
-
-