PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Catalog Number 400SMTSFT0408
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thrombus (2101)
Date of Event 02/13/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for evaluation.From the information provided, there is no indication that there was any device malfunction, nonconformance, or misuse that contributed to the reported event.Thrombosis is a known and anticipated complication with these types of procedures and is noted in the device labeling.Therefore, it was determined that the reported thrombus formation was an anticipated procedural complication.The product lot number was not provided; therefore, the manufacturing records could not be reviewed.This report is associated with mfr report numbers: 3005168196-2018-00562, 3005168196-2018-00563, 3005168196-2018-00565, 3005168196-2018-00566.The device was implanted into the patient.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the right middle cerebral artery (mca) using penumbra smart coils (smart coils).During the procedure, the physician successfully placed five smart coil and a non-penumbra stent.During the procedure there was thrombus formation, and therefore the hospital staff immediately administered thrombolytics intra-arterially.There was complete restoration of flow, and the patient demonstrated no new residual neurologic deficits from this event.This adverse event was resolved.The thrombus formation was adjudicated to have a probable relationship to the smart coil system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7339858
Report Number3005168196-2018-00564
Device Sequence Number13713473
Product Code HCG
UDI-Device Identifier00814548015088
UDI-Public00814548015088
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Catalogue Number400SMTSFT0408
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2018
Initial Report FDA Received Date03/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age51 YR
-
-