| Model Number |
ZXT225 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017 & (b)(6) 2018.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tecnis symfony toric multifocal intraocular lens (model zxt225, +16.5 diopter) was explanted in a secondary procedure from patient's left eye because of residual refractive error and patient complaining of blurry vision.Reportedly model zxt300, +18.0 diopter was used as replacement lens for the patient.No incision enlargement, no vitrectomy was performed, no other medical or surgical intervention was required.The patient was reported doing well when discharged.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation ¿ yes, returned to manufacturer on 3/15/2018.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|