MAUDE Adverse Event Report: COCHLEAR BONE ANCHORED SOLUTIONS AB BIA400 IMPLANT 4MM W ABUTMENT 10MM COCHLEAR BAHA CONNECT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR BONE ANCHORED SOLUTIONS AB BIA400 IMPLANT 4MM W ABUTMENT 10MM COCHLEAR BAHA CONNECT SYSTEM Back to Search Results
Model Number 93331
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Host-Tissue Reaction (1297)
Event Date 01/15/2018
Event Type  Injury  
Manufacturer Narrative

The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time. Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication. Additionally, there are no indications that a product failure has contributed to the reported issue. Adverse skin reactions around the baha abutment, ranging from slight redness to infected tissue, are common complications with baha implants, and can occur at any time following implantation. The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis. This event is added to this monitoring. This report is submitted on march 15, 2018. (b)(4).

 
Event Description

Per the clinic, the patient experienced skin overgrowth at abutment site; subsequently on (b)(6) 2018, the patient underwent skin revision surgery to excise skin during an abutment change. The implanted device remains.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameBIA400 IMPLANT 4MM W ABUTMENT 10MM
Type of DeviceCOCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
po box 82
mölnlycke, 435 2 2
SW 435 22
Manufacturer Contact
kristel kohne
1 university avenue
macquarie university, nsw 2109,
AS   2109,
MDR Report Key7340524
MDR Text Key102434725
Report Number6000034-2018-00543
Device Sequence Number1
Product Code LXB
Combination Product (Y/N)N
Reporter Country CodeIS
PMA/PMN NumberK121317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation HEALTH PROFESSIONAL
Type of Report Initial
Report Date 02/27/2018
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received03/15/2018
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? No
Device Operator LAY USER/PATIENT
Device MODEL Number93331
Device Catalogue Number93331
Device LOT NumberN/A
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer03/14/2018
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received02/27/2018
Was Device Evaluated By Manufacturer? No
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

Patient TREATMENT DATA
Date Received: 03/15/2018 Patient Sequence Number: 1
-
-