| Model Number |
HEWAA2810080810/1 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
02/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the complaint device history records indicated that the graft was processed and inspected according to established procedures and was therefore released following acceptable quality inspections and tests.Specifically, no anomaly was evidenced in the collagen coating records.The review of the water permeability testing records of products coated on the same day as the complaint device indicated values well within product specifications.Please note that water permeability testing is recognized by international standard for vascular grafts as the test to address the risk of blood leakage.(b)(4).One retention sample coated on the same period and under the same conditions as the involved device underwent water permeability testing.The test result indicated a value well within product specifications (< 5 ml/cm²/min).(b)(4).No conclusion can be drawn.However, the conducted investigation and testing performed would tend to indicate that the product was not defective.
|
| |
|
Event or Problem Description
|
|
It was reported that, on (b)(6) 2018, during a surgery for an aneurysmal disease, a serious bleeding was observed from the graft.The blood loss was about 5.5k cc and a transfusion was required during the surgery.Hemostatic agent (novoseven[sup]) and blood plasma cryoprecipitate were used to stop the bleeding.The graft remained implanted.No consequence for the patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The case has been submitted to the corporate medical officer for review, with two other similar cases reported together by the same physician (mdr # 1640201-2018-00008 & 164201-2018-00009).His medical opinion is reported as follows: "we have received three complaints regarding complex aortic surgeries performed at same institution in china by the surgeon.Very limited information has been collected regarding the preexisting conditions of the patients.In all cases a very significant blood loss from the totality of the vascular grafts, from 3.5 to 5.5 liters.For all cases hemostatic agents were required to stop the bleeding.The procedures were significantly delayed by the hemostatic management.Ultimately the blood loss was controlled and the graft remained implanted in all cases.Similarly, all patients had no clinically significant consequences due to the events described.The limited information provided don't allow a clinical assessment.However the magnitude of the blood loss was very significant.The cause of the bleeding was unknown.Despite the normal post-operative course, the patients could have experience very serious consequences that could have been life-threatening." the additional information does not modify the conclusion provided in the initial report.
|
| |
|
Event or Problem Description
|
|
The graft was used during a surgery with ecc (extracorporeal circulation) for an aortic aneurysm.The bleeding was observed from both main body and branches (including perfusion branch).
|
| |
|