Model Number METRIX AIR
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause : mlc-55 user had an inaccurate reference: competitor's meter: the end user is comparing results obtained from trividia's bgm system to the results from a competitor's bgm system note: manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
 
Event or Problem Description
Consumer reported complaint for high blood glucose results.Wife is calling on behalf of the customer.The customer is concerned with results obtained of 223mg/dl.The expected fasting blood glucose test result range is 80-250mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the dining room.During the call on (b)(6) 2018, a back to back blood test was not performed.The test strip lot manufacturer's expiration date is 10/31/2018 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly).Result 1:44mg/dl date: (b)(6) 2018 time:8:00pm control.Result 2:223mg/dl date: (b)(6) 2018 time:1:41pm fasting.Result 3:98mg/dl date: (b)(6) 2018 time:9:13am fasting.Result 4:120mg/dl date: (b)(6) 2018 time:5:52pm fasting.Result 5:235mg/dl date: (b)(6) 2018 time:3:17pm n/fasting.Customer says that at her husband's most recent doctor's appointment, his glucose level was checked using the true metrix meter and another meter.Customer says that on the true metrix, result was 235mg/dl and on the other meter, result was 185mg/dl.Customer says that lab results were also significantly lower than the true metrix and that her husband's most recent a1c level was 8.1%.Customer ran control test before calling and is still not comfortable with meter even though results were within range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7341224
Report Number1000113657-2018-00307
Device Sequence Number19200091
Product Code NBW
UDI-Device Identifier00021292009472
UDI-Public(01)00021292009472(17)181031(10)MU2461
Combination Product (Y/N)N
Initial Reporter StateMD
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 03/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMU2461
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2018
Initial Report FDA Received Date03/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-